Free-Breathing Water, Fat, ^{\star}$ and $ Field Mapping of the Liver Using Multi-Echo Radial FLASH and Regularized Model-based Reconstruction (MERLOT)

2021-01-07 | preprint. A publication with affiliation to the University of Göttingen.

Jump to: Cite & Linked | Documents & Media | Details | Version history

Cite this publication

​Free-Breathing Water, Fat, ^{\star}$ and $ Field Mapping of the Liver Using Multi-Echo Radial FLASH and Regularized Model-based Reconstruction (MERLOT)​
Tan, Z.; Rosenzweig, S.; Wang, X.; Scholand, N.; Holme, C. ; Blumenthal, M.& Voit, D. et al.​ (2021)

Documents & Media

License

GRO License GRO License

Details

Authors
Tan, Zhengguo; Rosenzweig, Sebastian; Wang, Xiaoqing; Scholand, Nick; Holme, Christian ; Blumenthal, Moritz; Voit, Dirk; Frahm, Jens ; Uecker, Martin 
Abstract
Purpose: To achieve free-breathing quantitative fat and $R_2^{\star}$ mapping of the liver using a generalized model-based iterative reconstruction, dubbed as MERLOT. Methods: For acquisition, we use a multi-echo radial FLASH sequence that acquires multiple echoes with different complementary radial spoke encodings. We investigate real-time single-slice and volumetric multi-echo radial FLASH acquisition. For the latter, the sampling scheme is extended to a volumetric stack-of-stars acquisition. Model-based reconstruction based on generalized nonlinear inversion is used to jointly estimate water, fat, $R_2^{\star}$, $B_0$ field inhomogeneity, and coil sensitivity maps from the multi-coil multi-echo radial spokes. Spatial smoothness regularization is applied onto the B 0 field and coil sensitivity maps, whereas joint sparsity regularization is employed for the other parameter maps. The method integrates calibration-less parallel imaging and compressed sensing and was implemented in BART. For the volumetric acquisition, the respiratory motion is resolved with self-gating using SSA-FARY. The quantitative accuracy of the proposed method was validated via numerical simulation, the NIST phantom, a water/fat phantom, and in in-vivo liver studies. Results: For real-time acquisition, the proposed model-based reconstruction allowed acquisition of dynamic liver fat fraction and $R_2^{\star}$ maps at a temporal resolution of 0.3 s per frame. For the volumetric acquisition, whole liver coverage could be achieved in under 2 minutes using the self-gated motion-resolved reconstruction. Conclusion: The proposed multi-echo radial sampling sequence achieves fast k -space coverage and is robust to motion. The proposed model-based reconstruction yields spatially and temporally resolved liver fat fraction, $R_2^{\star}$ and $B_0$ field maps at high undersampling factor and with volume coverage.
Issue Date
7-January-2021
Project
EXC 2067: Multiscale Bioimaging 
Working Group
RG Uecker 

Reference

Citations