There is no F in APC: Using physiological fluoride-free solutions for high throughput automated patch clamp experiments

2022 | journal article. A publication with affiliation to the University of Göttingen.

Jump to: Cite & Linked | Documents & Media | Details | Version history

Cite this publication

​There is no F in APC: Using physiological fluoride-free solutions for high throughput automated patch clamp experiments​
Rapedius, M.; Obergrussberger, A.; Humphries, E. S. A.; Scholz, S.; Rinke-Weiss, I.; Goetze, T. A. & Brinkwirth, N. et al.​ (2022) 
Frontiers in Molecular Neuroscience15 art. 982316​.​ DOI: https://doi.org/10.3389/fnmol.2022.982316 

Documents & Media

Data_sheet_1.pdf188.48 kBUnknownfnmol-15-982316.pdf5.79 MBUnknownfnmol-15-982316-g001.tif191.21 kBUnknownfnmol-15-982316-g002.tif176.46 kBUnknownfnmol-15-982316-g003.tif214.95 kBUnknownfnmol-15-982316-g004.tif221.18 kBUnknownfnmol-15-982316-g005.tif272.98 kBUnknownfnmol-15-982316-g006.tif234.65 kBUnknownfnmol-15-982316-g007.tif58.82 kBUnknown

License

Published Version

Attribution 4.0 CC BY 4.0

Details

Authors
Rapedius, Markus; Obergrussberger, Alison; Humphries, Edward S. A.; Scholz, Stephanie; Rinke-Weiss, Ilka; Goetze, Tom A.; Brinkwirth, Nina; Rotordam, Maria Giustina; Strassmaier, Tim; Randolph, Aaron; Fertig, Niels; Friis, Søren; Liutkute, Aiste; Seibertz, Fitzwilliam; Voigt, Niels 
Abstract
Fluoride has been used in the internal recording solution for manual and automated patch clamp experiments for decades because it helps to improve the seal resistance and promotes longer lasting recordings. In manual patch clamp, fluoride has been used to record voltage-gated Na (Na V ) channels where seal resistance and access resistance are critical for good voltage control. In automated patch clamp, suction is applied from underneath the patch clamp chip to attract a cell to the hole and obtain a good seal. Since the patch clamp aperture cannot be moved to improve the seal like the patch clamp pipette in manual patch clamp, automated patch clamp manufacturers use internal fluoride to improve the success rate for obtaining GΩ seals. However, internal fluoride can affect voltage-dependence of activation and inactivation, as well as affecting internal second messenger systems and therefore, it is desirable to have the option to perform experiments using physiological, fluoride-free internal solution. We have developed an approach for high throughput fluoride-free recordings on a 384-well based automated patch clamp system with success rates >40% for GΩ seals. We demonstrate this method using hERG expressed in HEK cells, as well as Na V 1.5, Na V 1.7, and K Ca 3.1 expressed in CHO cells. We describe the advantages and disadvantages of using fluoride and provide examples of where fluoride can be used, where caution should be exerted and where fluoride-free solutions provide an advantage over fluoride-containing solutions.
Issue Date
2022
Journal
Frontiers in Molecular Neuroscience 
Project
EXC 2067: Multiscale Bioimaging 
SFB 1002: Modulatorische Einheiten bei Herzinsuffizienz 
SFB 1002 | A13: Bedeutung einer gestörten zytosolischen Calciumpufferung bei der atrialen Arrhythmogenese bei Patienten mit Herzinsuffizienz (HF) 
Organization
Institut für Pharmakologie und Toxikologie ; Universitätsmedizin Göttingen ; Deutsches Zentrum für Herz-Kreislauf-Forschung e.V. 
Working Group
RG Voigt (Molecular Pharmacology) 