Cervical Disc Arthroplasties Fail to Maintain Physiological Kinematics Under Lateral Eccentric Loads

2023 | journal article. A publication with affiliation to the University of Göttingen.

Jump to: Cite & Linked | Documents & Media | Details | Version history

Cite this publication

​Cervical Disc Arthroplasties Fail to Maintain Physiological Kinematics Under Lateral Eccentric Loads​
Roch, P. J.; Hemprich, C.; Klockner, F.; Meier, M.-P.; Jäckle, K.; Lehmann, W. & Wachowski, M. M. et al.​ (2023) 
Global Spine Journal, art. 21925682231193642​.​ DOI: https://doi.org/10.1177/21925682231193642 

Documents & Media

License

GRO License GRO License

Details

Authors
Roch, Paul Jonathan; Hemprich, Constantin; Klockner, Friederike; Meier, Marc-Pascal; Jäckle, Katharina; Lehmann, Wolfgang; Wachowski, Martin Michael; Weiser, Lukas
Abstract
Study design In vitro human cadaveric biomechanical analysis. Objectives Optimization of prostheses for cervical disc arthroplasties (CDA) reduces the risk of complications. The instantaneous helical axis (IHA) is a superior parameter for examining the kinematics of functional spinal units. There is no comprehensive study about the IHA after CDA considering all 3 motion dimensions. Methods Ten human functional spinal units C4-5 (83.2 ± 7.9 yrs.) were examined with an established measuring apparatus in intact conditions (IC), and after CDA, with 2 different types of prostheses during axial rotation, lateral bending, and flexion/extension. Eccentric preloads simulated strains. The IHA orientation and its position at the point of rest (IHA 0 -position) were analyzed. Results The results confirmed the existing data for IHA in IC. Lateral preloads showed structural alterations of kinematics after CDA: During axial rotation and lateral bending, the shift of the IHA 0 -position was corresponding with the lateral preloads’ applied site in IC, while after CDAs, it was vice versa. During lateral bending, the lateral IHA orientation was inclined, corresponding with the lateral preloads’ applied site in the IC and oppositely after the CDAs. During flexion/extension, the lateral IHA orientation was nearly vertical in the IC, while after CDA, it inclined, corresponding with the lateral preloads’ applied site. The axial IHA orientation rotated to the lateral preloads’ corresponding site in the IC; after CDA, it was vice versa. Conclusion Both CDAs failed to maintain physiological IHA characteristics under lateral preloads, revealing a new aspect for improving prostheses’ design and optimizing their kinematics.
Issue Date
2023
Journal
Global Spine Journal 
ISSN
2192-5682
eISSN
2192-5690
Language
English

Reference

Citations