In-line monitoring of electrolytes and urea during continuous renal replacement therapy

2020 | journal article. A publication with affiliation to the University of Göttingen.

Jump to: Cite & Linked | Documents & Media | Details | Version history

Cite this publication

​In-line monitoring of electrolytes and urea during continuous renal replacement therapy​
Berger, M.; Sehlmeyer, M.; Sellering, F.; Röhrich, H.; Perl, T.; Mansour, H. & Baasner, D. et al.​ (2020) 
Journal of Sensors and Sensor Systems9(2) pp. 251​-262​.​ DOI: https://doi.org/10.5194/jsss-9-251-2020 

Documents & Media

License

GRO License GRO License

Details

Authors
Berger, Marc; Sehlmeyer, Merle; Sellering, Flora; Röhrich, Hannes; Perl, Thorsten; Mansour, Hussam; Baasner, Daniel; Zimmermann, Stefan
Abstract
Abstract. The individualization of dialysis treatment using a customized dialysate composition usually requires a continuous measurement of electrolytes and urea in blood. The current practices are spot measurements of blood samples either with blood gas analyzers or in the laboratory, involving considerable personnel effort. Furthermore, the measured values are time delayed and not available in a continuous fashion. In this paper we investigate an in-line concept for continuous monitoring of important blood parameters such as sodium, potassium, calcium and urea concentrations in blood serum using ion-selective electrodes. This concept is evaluated in a preclinical study with human packed red blood cells as a test medium over a period of 7 h. It has been shown that the electrolytes can be well monitored. In addition, we present first measurements with ion-sensitive field-effect transistors in a miniaturized sensor assembly. Therefore, new low-cost electronics for such ion-sensitive field-effect transistors have been developed.
Issue Date
2020
Journal
Journal of Sensors and Sensor Systems 
eISSN
2194-878X
Language
English

Reference

Citations