​Indium K α radiation from a MetalJet X-ray source: comparison of the Eiger2 CdTe and Photon III detectors​
Ruth, P. N.; Graw, N.; Ernemann, T.; Herbst-Irmer, R. & Stalke, D.​ (2023) 
Journal of Applied Crystallography56(5) pp. 1322​-1329​.​ DOI: https://doi.org/10.1107/S1600576723007215 

Documents & Media

License

GRO License GRO License

Details

Authors
Ruth, Paul Niklas; Graw, Nico; Ernemann, Tobias; Herbst-Irmer, Regine; Stalke, Dietmar
Abstract
The MetalJet source makes available new K α radiation wavelengths for use in X-ray diffraction experiments. The purpose of this paper is to demonstrate the application of indium K α radiation in independent-atom model refinement, as well as approaches using aspherical atomic form factors. The results vary greatly depending on the detector employed, as the energy cut-off of the Eiger2 CdTe provides a solution to a unique energy contamination problem of the MetalJet In radiation, which the Photon III detector cannot provide.