INFLUENCE OF INTRINSIC NOISE GENERATED BY A THERMOTESTING DEVICE ON THERMAL SENSORY DETECTION AND THERMAL PAIN DETECTION THRESHOLDS

2009 | journal article. A publication with affiliation to the University of Göttingen.

Jump to: Cite & Linked | Documents & Media | Details | Version history

Cite this publication

​INFLUENCE OF INTRINSIC NOISE GENERATED BY A THERMOTESTING DEVICE ON THERMAL SENSORY DETECTION AND THERMAL PAIN DETECTION THRESHOLDS​
Pavlakovic, G.; Zuechner, K.; Zapf, A.; Bachmann, C. G.; Graf, B. M.; Crozier, T. A. & Pavlakovic, H.​ (2009) 
Muscle & Nerve40(2) pp. 257​-263​.​ DOI: https://doi.org/10.1002/mus.21270 

Documents & Media

License

GRO License GRO License

Details

Authors
Pavlakovic, Goran; Zuechner, Klaus; Zapf, Antonia; Bachmann, Cornelius G.; Graf, Bernhard Martin; Crozier, Thomas A.; Pavlakovic, Helena
Abstract
Various factors can influence thermal perception threshold measurements and contribute significantly to unwanted variability of the tests. To minimize this variability, testing should be performed under strictly controlled conditions. Identifying the factors that increase the variability and eliminating their influence should increase reliability and reproducibility. Currently available thermotesting devices use a water-cooling system that generates a continuous noise of approximately 60 dB. In order to analyze whether this noise could influence the thermal threshold measurements we compared the thresholds obtained with a silent thermotesting device to those obtained with a commercially available device. The subjects were tested with one randomly chosen device on 1 day and with the other device 7 days later. At each session, heat, heat pain, cold, and cold pain thresholds were determined with three measurements. Bland-Altman analysis was used to assess agreement in measurements obtained with different devices and it was shown that the intersubject variability of the thresholds obtained with the two devices was comparable for all four thresholds tested. In contrast, the intrasubject variability of the thresholds for heat, heat pain, and cold pain detection was significantly lower with the silent device. Our results show that thermal sensory thresholds measured with the two devices are comparable. However, our data suggest that, for studies with repeated measurements on the same subjects, a silent thermotesting device may allow detection of smaller differences in the treatment effects and/or may permit the use of a smaller number of tested subjects. Muscle Nerve 40: 257-263, 2009
Issue Date
2009
Status
published
Publisher
John Wiley & Sons Inc
Journal
Muscle & Nerve 
ISSN
0148-639X
Sponsor
B. Braun Foundation

Reference

Citations